Installation of TLC Visualizer camera¶
To use the Baumer camera with your computer, you have to install the Baumer-specific driver for your FireWire card (OHCI driver) as well as the Baumer driver for the Camera (Camera driver) itself.
Warning
If the driver installed is not the correct one, CAMAG cannot guarantee correct functioning of the camera.
OHCI driver installation¶
Open Device Manager, check for IEEE1394 host controller, right click on the entry representing the controller connected to your camera and select Update Driver Software
Do not allow Windows to search automatically for drivers. Manually search on the visionCATS CD for the correct driver
Choose Choose from a list of driver
Make sure the compatible hardware checkbox is active
Use Have Disc to search the visionCATS disc
Choose your format (x64 for 64 bit, x86 for 32bit)
Choose OHCI
Choose the respective chip set
Warning
If the chip set your computer is using is not listed, do not install it. Change the FireWire card to a product with a supported chip set
Camera driver installation¶
Connect the camera with the computer
Do not accept any driver installation yet
Open Device Manager, check for Baumer_Optronic&&Camera, right click on this device and select Update Driver Software.
Do not allow Windows to search automatically for drivers. Manually search on the visionCATS CD for the correct driver
Choose Choose from a list of driver
Browse the visionCATS disc for the correct driver and install it
Restart your pc
If the Visualizer is not recognized by visionCATS check the following:
Is the instrument switched on?
Is the camera (FireWire) and the Visualizer (serial) connected with the pc?
Is the green light at the camera on?
Is the firmware of camera and Visualizer up to date?
Configuration for Windows 10¶
The TLC Visualizer camera has some compatibility issues with Windows 10. For avoiding them you need to configure the PC:
Open Power Options
Click on Choose what the power button does on the left menu
If needed, click on Change settings that are currently unavailable
Uncheck Turn on fast startup
Important
If you get a blue screen when capturing image with TLC Visualizer, fast startup may again be activated. Please do the above configuration.